Merit PreludeSync Radial Compression Device Mar 6, 2017 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer Merit Medical Systems Inc. 1600 West Merit Parkway South Jordan 84095 Utah UNITED STATES Merit PreludeSync Radial Compression Device SRB24AC
Wrist band fiber particulate was identified on the sealing surface inside the valve, preventing a complete seal of the inflation valve resulting in a slow balloon air leak.