Pentax Video Duodenoscope

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Hoya Corporation
6-10-1 NISHI-SHINJUKU
Shinjuku-ku
160-0023
JAPAN
Pentax Video DuodenoscopeED-3490TK

Reason

During interactive discussions with US FDA concerning a 510(k) additional information ("AI") request for the ED-3490TK duodenoscopes, the FDA requested that Pentax perform a root cause analysis to determine how soiling occurred on the surface of the suction cylinder and under the distal cap during testing. FDA also requested that Pentax identify risk mitigation strategies to address the manufacturing process, possible damage to the duodenoscope with use, and any other aspects of risk mitigation. As a result of the communications with FDA, and at the suggestion of FDA, it was decided that Pentax would issue a customer safety communication to advise customers who use ED-3490TKS about a potential issue associated with the distal cap and offer recommendations intended to reduce the potential risk for contamination and subsequent patient infection. Specifically, the user communication states, "during manufacturing of ED-3490TK duodenoscopes, silicone adhesive is applied to the distal tip prior to affixing the distal cap. In some instances, cracks or gaps may form in the adhesive that may be vulnerable to fluid ingress and soiling." the letter reminds users of the importance of using the ed-3490tk in accordance with its current labeling and assuring that all reprocessing personnel are knowledgeable and thoroughly trained on the current IFU for manual reprocessing of these devices. Additionally, the letter recommends that users "immediately remove from use any ed-3490tk duodenoscope that shows visible signs of wear or physical damage" and states that Pentax will provide on-site inspections to assess duodenoscope functionality, with particular attention to the integrity of the distal cap, and provide in-house servicing as needed.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime