Health Products

1,25-Dihydroxy Vitamin D Ria

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
J400071,25-Dihydroxy Vitamin D Ria
Manufacturer
Immunodiagnostic Systems Limited
10 Didcot Way, Boldon Business Park,
Boldon, Tyne and Wear,
NE35 9PD
UNITED KINGDOM
AA-54F1

Reason

In a small number, approximately 6%, of 1,25-dihydroxy vitamin d ria kits containing calibrators D lot 34649 and E lot 34650, similar CPM values for both calibrators are obtained. These abnormal results lead to an incorrect calibration curve fit.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime