Confirmation of a screen printing error mentionned on the strips of the impacted lots. It is written that 5 microliters must be pipetted and transferred to the wells "d" and "e". This information is wrong, the right volume to transfer is 25 microliters as specified in the instruction for use (ref. 20584, version e). This screen printing error could lead to a wrong protocol and pipetting 5 microliters instead of 25 microliters. When following the wrong protocol, there is not issue for the Carbapenamase positive strains all correctly identified as positive, but there is a potential risk of false positive result for Carbapenamase negative strains.