Health Products

Dimension Vista Systems Assays (2017-04-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsF. Dimension Vista System Loci Estradiol (E2) Reagent
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
10489099
All lotsD. Dimension Vista System-Loci Digoxin (Digxn) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
10488927
All lotsE. Dimension Vista System-Loci Prolactin (Prl) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
10488398
All lotsB. Dimension Vista System-Ferritin (Ferr) Assay (Flex)
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
K6440
All lotsC. Dimension Exl System-(Troponin) Tni Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
10464525
All lotsA. Dimension Vista System-Thyroid Stimulating Hormone (Tsh) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
K6412

Introduction

Affected Products
  1. Dimension Vista System-Thyroid Stimulating Hormone (Tsh) Assay
  2. Dimension Vista System-Ferritin (Ferr) Assay (Flex)
  3. Dimension Exl System-(Troponin) Tni Assay
  4. Dimension Vista System-Loci Digoxin (Digxn) Assay
  5. Dimension Vista System-Loci Prolactin (Prl) Assay
  6. Dimension Vista System Loci Estradiol (E2) Reagent

Reason

Siemens Healthcare Diagnostics has confirmed that the concentrations for biotin listed in the non- interfering substances section of the current dimension and dimension vista instructions for use for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime