Health Products

ProBP 2400

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Microlife Corporation
9F, 431 Ruiguang Road, Nei-Hu
Taipei
11492
TAIWAN, PROVINCE OF CHINA
ProBP 2400901096

Reason

The manufacturer (Microlife Corp.) incorrectly installed a component on the printed circuit boards in (2) lots of the ProBP 2400 during rework in the manufacturing process. As a result of the defective electrical component in the battery charging circuit, when the ProBP 2400 is connected to the external power supply/battery charger, a potential over-voltage battery charging condition may occur. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic (ABS) battery door and other adjacent plastic and foam device components.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime