Cobas Chemistry Analyzers (2017-03-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
15F6-07 to 17M8-10
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
F. Cobas 8000-C702 Chemistry Analyzer6473245001
15A6-01 to 16D9-10
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
D. 559756  Cobas 8000-C502 Chemistry Analyzer5964067001
15E6-01 to 17H7-10
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
E. Cobas 8000-C701 Chemistry Analyzer5641489001
15D0-01 to 16D8-20
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
B. Cobas 4000 - Cobas C 3114826876001
15D5-01 to 17L8-10
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
C. ISE1800 Additional Module For Cobas 80005964075001
15P1-01 to 16Y4-19
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. Cobas 6000 - C501 Chemistry Analyzer4745914001

Introduction

Affected Products

A. Cobas 6000 - C501 Chemistry Analyzer
B. Cobas 4000 - Cobas C 311
C. ISE1800 Additional Module For Cobas 8000
D. 559756  Cobas 8000-C502 Chemistry Analyzer
E. Cobas 8000-C701 Chemistry Analyzer
F. Cobas 8000-C702 Chemistry Analyzer

Reason

Following a production change of connector types for the sample probe plug, it was confirmed by the manufacturer that in very rare cases, a disturbance of the sample liquid level detection (LLD) may occur. This may cause the affected sample probe to dip into the sample material deeper than intended, causing the affected sample probe to be washed inadequately and lead to a potential carryover.