Health Products

Omega Plus Standard Lag Screw

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
V06269
Manufacturer
Stryker GMBH
Bohnackerweg 1,
Selzach, Solothurn
2545
SWITZERLAND
Omega Plus Standard Lag Screw3362-5-105

Reason

Stryker received a complaint regarding a Standard Lag Screw Omega 110mm length. It was reported that the lag screw was not completely cannulated and the guide wire could not go through the cannula of the screw. No harm reported. The investigation revealed that the obstruction is related to a metal burr in the cannula of the screw which was not removed during manufacturing. The non-conformance is limited to 2 batches.