A. Advia Centaur System - Insulin Assay
B. Advia Centaur CP System - Insulin Assay
Reason
Siemens has identified that current in-date lots of Advia Centaur Systems Insulin Assay, recover approximately 40% higher than the world health organization (WHO) 1st IRP 66/304 standardization based on slope values. The slope value was 1.40 on the Advia Centaur and 1.42 on the Advia Centaur CP. Corrected (re-standardized) lots will be available by May 2017.