Siemens Healthcare Diagnostics has confirmed a high rate of dilution check relative bias failures of greater than 2% when using V-Lyte dilution check solution lots KC6D01, KC6F01, or KC6J01. Newer lots of dilution check starting with KC7A01 are not affected by this issue.