V-Lyte Integrated Multisensor Technology (Imt)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
V-Lyte Integrated Multisensor Technology (Imt)K800A

Reason

Siemens Healthcare Diagnostics has confirmed an increase in complaints regarding elevated urine sodium quality control (QC) exceeding acceptable QC ranges with the Dimension Vista IMT System. Siemens observed a greater urine sodium drift with urine QC than urine patient samples. Not all IMT sensors exhibit the described urine sodium QC drift. Current lots and all future lots are potentially affected until this issue is resolved.