Zimmer Packaged Implants (2017-05-09) May 9, 2017 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES F. Zimmer Natural Nail System Femoral Nail 47-2492-460-14 All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES D. Itst Femoral Nails 00-2257-300-04 All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES E. Zimmer Natural Nail System Tibial Nails 47249538014 All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES B. Versa Fx Femoral Fixation Tube Plates 00-1193-090-16 All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES C. M/Dn Intramedullary Fixation System 00-2240-010-48 All lots expiring before May 2027 Manufacturer Zimmer Inc. 1800 West Center Street Warsaw 46580 UNITED STATES A. Versa Fx Femoral Fixation Tube Plates 00-1193-130-16
Zimmer is initiating a field action for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the two configurations failed simulated distribution and shipping testing. The devices impacted are generally the heaviest outlier sizes within the respective product family.