Health Products

Twin Coil

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial number: 1023
Manufacturer
Tonica Elektronik A/S
Lucernemarken 15
Farum
3520
DENMARK
Twin Coil9016E0691

Reason

The coil housing broke apart due to a breakdown in the coil winding element inside the housing. The breakdown occurred near the outlet from the outer winding. The breakdown had been initiated from mechanical stress near the outlet. Such breakdown can start in the wires that are bended or twisted. During stimulation some mechanical displacement/movements will occur inside the litze wire and between the windings.