Infinity Delta Patient Monitor (2017-03-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not ApplicableB. Infinity Delta XL Patient Monitor
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES
MS18596
5396182958A. Infinity Delta Patient Monitor
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES
MS18597

Introduction

Affected Products

A. Infinity Delta Patient Monitor
B. Infinity Delta XL Patient Monitor

Reason

Draeger was informed about a case in which the affected units running VF10.0 software did not generate a set low O2 alarm although the O2 level that was measured by the Draeger Scio gas measurement module was below the adjusted alarm limit. Draeger analysis revealed a failure within the VF10.0 software. Earlier software versions are not affected.