Health Products

Spectra Optia Apheresis System - Main Unit

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1P04028 to 1P04392
Manufacturer
Terumo BCT Inc.
10811 W. Collins Avenue
Lakewood
Colorado
UNITED STATES
Spectra Optia Apheresis System - Main Unit61000

Reason

A power filter cable, a component used in the internal electrical system, had a visible defect which caused the unit to lose power. Further evaluation determined that this defect may be present in other electrical cables which could cause a similar non-recoverable power failure. The cables have been isolated to specific manufacturing lots