Health Products

Trima Accel Automated Blood Collection System, VE

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1T07708-1T08066
Manufacturer
Terumo BCT Inc.
10811 W. Collins Avenue
Lakewood
Colorado
UNITED STATES
Trima Accel Automated Blood Collection System, VE81000

Reason

A Trima Accel device demonstrated a non-recoverable power failure. A subsequent investigation of the failure determined that a power filter cable, a component used in the internal electrical system, had a visible defect which caused the unit to lose power. Further evaluation determined that this defect may be present in other electrical cables which could cause a similar non-recoverable power failure.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime