Somatom Force

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version du logiciel VA50A
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
Somatom Force10742326

Reason

An addendum to the instructions for use is being provided for Somatom Force Systems with software version VA50A to inform users about a potential malfunction for turbo flash scans with very short scan ranges. When selecting the turbo flash mode in combination with a spiral scan length of less than 27 mm the image reconstruction will possibly fail for this scan. As a work around the addendum suggests for the turbo flash mode to always select a spiral scan length higher than 27 mm or to switch to another mode instead.