Coulter PrepPlus and PrepPlus 2 (2017-05-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2.04
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
B. Coulter PrepPlus 2378600
1.00
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
A. Coulter PrepPlus286600

Introduction

Affected Products
  • A. Coulter PrepPlus
  • B. Coulter PrepPlus 2

Reason

When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a flow cytometer, there is imprecision in the absolute counts on only the first tube in a run when compared to the rest of the tubes in the same run. The issue may generate erroneous results for the reported parameters assayed on the sample in the first tube of the flow count worklist.