Health Products

Anti-Saccharomyces Cerevisiae Antibodies (Asca) Iga Test

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
16325A
Manufacturer
Bio-Rad Laboratories Clinical Diagnostics Group
4000 Alfred Nobel Drive
Hercules
94547
California
UNITED STATES
Anti-Saccharomyces Cerevisiae Antibodies (Asca) Iga Test4252320

Reason

It has been observed recently that the values of some samples tested with the autoimmune EIA ASCA IgA test decreased significantly when compared to the values of the same set of samples tested with another lot of autoimmune EIA ASCA IgA test. This observation was confirmed by subsequent testing with a different set of samples. The observed decrease in sample value presents a potential for increase in false negative results.