Health Products

Immulite Anti-Tg Ab Systems (2017-06-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
102
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
B. Immulite 1000 System - Anti-Tg AbLTGCVM
102
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
A. Immulite System - Anti-Tg AbLTGCVM

Introduction

Affected Products

A. Immulite System - Anti-Tg Ab
B. Immulite 1000 System - Anti-Tg Ab

Reason

Siemens has confirmed that Immulite Systems Anti-TG Ab calibration verification material (CVM), lot 102, levels CVM2 through CVM4, catalog LTGCVM, is recovering high outside of the published +/- 2SD guideline ranges in the IFU. This issue applies to use with the Immulite 2000 and Immulite 2000 XPI Anti-TG AB assay only.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime