Ak 98 Dialysis Machine

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial numbers 13320 and lower
Manufacturer
Gambro Lundia AB
MAGISTRATSVAGEN 16, P.O. BOX 10101,
Lund
22010
SWEDEN
Ak 98 Dialysis Machine115251

Reason

Baxter Corporation is issuing an urgent device correction for the AK 98 hemodialysis machines listed below. Baxter has identified that an incorrect cable was installed in the affected AK 98 machines. If the user connects the machines to the Ethernet network, a leakage of currents higher than specified values may occur. These units do not have, and are not intended to have, Ethernet functionality. Baxter will be correcting the affected units by providing a cap for the Ethernet connectors to block the connection.