Health Products

CH-200 Ceiling Tube Support

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
016M46108
Manufacturer
Shimadzu Corporation
1, Nishinokyo-Kuwabaracho
Nakagyo-Ku
604-8511
JAPAN
CH-200 Ceiling Tube SupportCH-200

Introduction

Affected products

CH-200 Ceiling Tube Support

Reason

The x-ray tube assembly in this device is mounted on a tube mounting flange, which is part of the tube holding shaft. It was found that, extremely infrequently, cracks may occur over time on the tube holding shaft near the base of the tube mounting flange. If these cracks occur, and then increase in size, the tube mounting flange may separate from the tube holding shaft.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime