Leica Asp6025 Tissue Processor

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
314Leica Asp6025 Tissue Processor
Manufacturer
LEICA BIOSYSTEMS NUSSLOCH GMBH
HEIDELBERGER STREET 17-19
NUSSLOCH
69226
GERMANY
ASP6025

Reason

The affected devices have incorrect labelling with regard to the specified voltage for the alarm connectors on the rear side of the instrument. The current incorrect labels specify 125VA.C as maximum voltage. It has been identified, that if someone connects such 125VA.C to the 6.3mm stereo jack, there would be hazardous voltage accessible on the plug which could lead in the worst case to a serious electrical incident. In addition, the current specification in the instruction for use (IFU)has an incorrect labelling with regard to the specified voltage for the alarm connectors. The translated IFU’s and the "online help" of the instruments also have an incorrect labelling with regard to the specified voltage of the alarm connectors.