Health Products

CS300, CS100 and CS100i Intra Aortic Balloon Pumps (2017-06-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerB. CS100 Intra Aortic Balloon Pump
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
0998-00-3013-XX
More than 10 numbers, contact manufacturerC. CS100i Intra Aortic Balloon Pump
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
0998-00-0446H53
More than 10 numbers, contact manufacturerA. CS300 Intra Aortic Balloon Pump
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
0998-00-3023-XX

Introduction

Affected Products

A. CS300 Intra Aortic Balloon Pump
B. CS100 Intra Aortic Balloon Pump
C. CS100i Intra Aortic Balloon Pump

Reason

Maquet/Getinge has received a complaint involving a CS300 IABP that did not pump due to an electrical test failure code #58 (power up vent tests fail), maintenance code #3, and an autofill failure, which has been associated to a patient death due to the failure of the device to initiate therapy. An electrical test failure code #58 is caused by a solenoid valve requiring more power than the solenoid driver board can deliver to open the valve. The lack of power prevents the coil from moving the plunger causing the valve not to open. This is recognized by the system as an electrical test failure code #58 upon powering on the IAB. Units distributed after December 11, 2013 are not affected by the field correction.