Health Products

Dlp Pressure Display Set and Antegrade/Retrograde Stopcock Adapter (2017-06-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2016030813
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
B. Antegrade/Retrograde Stopcock Adapter13001
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
A. Dlp Pressure Display Set, Disposable61000

Introduction

Affected Products

A. Dlp Pressure Display Set, Disposable
B. Antegrade/Retrograde Stopcock Adapter

Reason

During the manufacturing process, Medtronic identified small pinholes, approximately 1 mm or less in diameter, in a single packaging configuration of sterile pouches. Based on internal investigation, all products in this packaging configuration manufactured since March 1, 2016 are potentially affected. The result is a potential sterile barrier breach.