Manufacturer
Elekta Business Area Software Systems IMPAC Medical Systems, Inc.
MosaiqMOSAIQ

Reason

A table shift in the vertical direction was inverted during treatment. An inverse shift will result in the patient being out of position by double the shift magnitude. A mac file is instrumental to the correct operation, if values are incorrectly input, inverse shifts may occur.