Due to a pressure increase after the cylinder switch-over, there are 2 possible issues identified:
1. The cylinder status icon can change and erroneously indicate remaining gas in an empty NO(nitric oxide) gas cylinder.
2. The alarm "no cylinder 1 missing (or empty)" or "no cylinder 2 missing (or empty)" can be deactivated. If the empty no cylinder is not replaced, both no cylinders can run empty and cause a stop in no delivery. There have been no reports of patient injury as a result of this issue.