B. CS300 Intra Aortic Balloon Pump System - Main Unit
0998-00-3023-53
S804785-B1
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
C. System 98XT Intra Aortic Balloon Pump
0998-00-0479-XX
S803492-B0
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
A. System 98 (LABP) Intra-Aortic Balloon Pump
0998-00-0446-XX
Introduction
Affected Products
System 98 (LABP) Intra-Aortic Balloon Pump
CS300 Intra Aortic Balloon Pump System - Main Unit
System 98XT Intra Aortic Balloon Pump
CS100 Intra Aortic Balloon Pump
Reason
Manufacturer has determined that condensation in the drain line tubing may lead to a false blood detection alarm activating when no blood is present. Also liquid spills, such as saline, can create bridges of resistance between the circuit components, causing the circuit not to function as intended. This can impact initiation or continuation of therapy.