Health Products

Datascope Corp. Intra Aortic Balloon Pump (2017-07-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
D. CS100 Intra Aortic Balloon Pump0998-00-3013-XX
More than 10 numbers, contact manufacturer
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
B. CS300 Intra Aortic Balloon Pump System - Main Unit0998-00-3023-53
S804785-B1
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
C. System 98XT Intra Aortic Balloon Pump0998-00-0479-XX
S803492-B0
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
A. System 98 (LABP) Intra-Aortic Balloon Pump0998-00-0446-XX

Introduction

Affected Products
  1. System 98 (LABP) Intra-Aortic Balloon Pump
  2. CS300 Intra Aortic Balloon Pump System - Main Unit
  3. System 98XT Intra Aortic Balloon Pump
  4. CS100 Intra Aortic Balloon Pump

Reason

Manufacturer has determined that condensation in the drain line tubing may lead to a false blood detection alarm activating when no blood is present. Also liquid spills, such as saline, can create bridges of resistance between the circuit components, causing the circuit not to function as intended. This can impact initiation or continuation of therapy.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime