Health Products

Mg2

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Mg206407358190

Reason

The Magnesium Gen 2 Assay (cat # 06407358190) may have low quality control recovery and frequent failed calibrations upon performing 2 points calibration when used on the Cobas C701-702 Analyzers after 7 days on board.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime