Health Products

Architect Free T4 Reagent

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
LISNAMUCK, LONGFORD
IRELAND
Architect Free T4 Reagent7K65-29

Reason

Recent data generated by Abbott have determined that the measuring interval is 0.40 ng/dl to 5.00 ng/dl. Measuring interval is defined as the range of values in ng/dl which meets the limits of acceptable performance for both imprecision and linearity. The current measuring interval for list numbers 7K65-29, 7K65-34 and 7K65-39 is 0.40 ng/dl to 6.00 ng/dl. A new FT4 assay file and updated package insert will be created reflecting the new measuring interval of 0.40 ng/dl to 5.00 ng/dl to be used with list numbers 7K65-29, 7K65-34 and 7K65-39.