Medtronic Spinous Process Clamp

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Navigation Inc.
826 Coal Creek Circle
Louisville
80027
Colorado
UNITED STATES
Medtronic Spinous Process Clamp9734715

Reason

Medtronic has become aware of instances where spine clamps have been damaged when forced open beyond their intended limits during use and will subsequently not open once attached to a patient's spinous process. When the spine clamp is forced open beyond its intended limits, a component (captive washer) may break off. The washer could then be inadvertently left behind in a patient's body if the breakage occurs during the procedure. If the washer is missing from the device, the spine clamp cannot be re-opened after placement on the spinous process.  The unintended removal of spinous process can lead to damage of adjacent vertebra and cause premature degradation.

This notification is intended to provide instructions to help healthcare practitioners identify whether a spine clamp has been damaged prior to use.