VerSys Beaded Hip Stem Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper and Zimmer Segmental System Distal Femoral Component (2017-08-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
D. Zimmer Segmental System Distal Femoral Component Size B00-5850-012-01
More than 100 numbers, contact manufacturer
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
E. Zimmer Segmental System Distal Femoral Component Size C00-5850-013-01
More than 10 numbers, contact manufacturer
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
B. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Extended Neck Offset Std Body00-7843-018-26
More than 10 numbers, contact manufacturer
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
C. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Extended Neck Offset Lm Body00-7843-017-46
More than 100 numbers, contact manufacturer
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
A. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Standard Neck Offset00784301406

Introduction

Affected Products

A. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Standard Neck Offset
B. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Extended Neck Offset Std Body
C. Versys Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper Extended Neck Offset Lm Body
D. Zimmer Segmental System Distal Femoral Component Size B
E. Zimmer Segmental System Distal Femoral Component Size C

Reason

Zimmer Biomet is recalling selected lots of VerSys Beaded Hip Stem Femoral Stem Beaded Fullcoat Collared 12/14 Neck Taper and Zimmer Segmental System Distal Femoral Component. Affected products were packaged in a previous configuration that was not tested for products weighing over 487 grams. If a product exceeds the weight tested, there is an increased likelihood of compromising the sterile barrier integrity.