Health Products

Oncentra External Beam/Oncentra Brachy 4.5

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 4.5
Manufacturer
Nucletron B.V.
Waardgelder 1
Veenedaal
3905 TH
NETHERLANDS
Oncentra External Beam/Oncentra Brachy 4.5170.730

Reason

The issue is caused by the measured source paths for the following applicators: ring applicator sets, ring CT/MR applicator sets, interstitial CT/MR rings, Vienna CT/MR rings and advanced gynecological applicator - Venezia with lunar-shaped ovoids, which have a source step size of 2.5 mm for the Microselectron Afterloader. If you use such an applicator model to create a plan, while the default step size of the Afterloader is 5.0, the step size in the ring or lunar-shaped ovoids will be incorrect. They will be shown as 2.5 mm, while the Afterloader will deliver at 5.0 mm if the error is not detected during plan approval.