Medtronic (Unomedical) Infusion Sets (2017-09-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Unomedical A/S
Aaholmvej 1-3
Osted
DENMARK
D. MIO INFUSION SETSMMT-921600
More than 10 numbers, contact manufacturer
Manufacturer
Unomedical A/S
Aaholmvej 1-3
Osted
DENMARK
E. MINIMED MIO 30MMT-905600
More than 10 numbers, contact manufacturer
Manufacturer
Unomedical A/S
Aaholmvej 1-3
Osted
DENMARK
B. PARADIGM SILHOUETTE INFUSION SETSMMT-368600
More than 10 numbers, contact manufacturer
Manufacturer
Unomedical A/S
Aaholmvej 1-3
Osted
DENMARK
C. PARADIGM QUICK-SET INFUSION SETSMMT-386600
More than 10 numbers, contact manufacturer
Manufacturer
Unomedical A/S
Aaholmvej 1-3
Osted
DENMARK
A. PARADIGM SURE-T INFUSION SETMMT-862

Introduction

Affected products
  1. PARADIGM SURE-T INFUSION SET
  2. PARADIGM SILHOUETTE INFUSION SETS
  3. PARADIGM QUICK-SET INFUSION SETS
  4. MIO INFUSION SETS
  5. MINIMED MIO 30

Reason

Medtronic has become aware of recent reports of potential over-delivery of insulin shortly after an infusion set change. The reported incidence rate requiring medical assistance related to this issue is less than 1 in every 2 million infusion sets. Over-delivery of insulin can cause hypoglycemia and in extreme cases, death. Medtronic has received reports of hypoglycemia requiring medical intervention potentially related to this issue.

Manufacturer investigation has shown this can be caused by fluid blocking the infusion set membrane during the priming/fill-tubing process. A membrane blocked by fluid may occur if insulin, alcohol, or water is spilled on top of the insulin reservoir, which then could prevent the infusion set from working properly. Infusion sets currently being shipped by Medtronic contain a new and enhanced membrane material that significantly reduces this risk.