Health Products

Amia Automated Peritoneal Dialysis Set With Cassette (2017-09-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
H16H04114Amia Automated Peritoneal Dialysis Set With Cassette
Manufacturer
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield
60015
Illinois
UNITED STATES
5C5479

Introduction

Affected Products

A. Amia Automated Peritoneal Dialysis Set With Cassette

Reason

Baxter Corporation is issuing a product recall for certain lots of the Amia Automated Peritoneal Dialysis Set with cassette in response to increased customer complaints received for missing red line, patient slow flow, solution slow flow, and inadequate drain alerts. Increased frequency of these alerts could lead to insufficient draining of waste solution, which can be a contributing factor in fluid overload.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime