Health Products

Persona The Personalized Knee System Tibila Sizing Plate (2017-08-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
63535352
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
B. Persona The Personalized Knee System Natural - Tibia Cemmented 5 Degree Stemmed Provisional42-5321-083-01
63531869
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
A. Persona The Personalized Knee System Tibila Sizing Plate42-5399-083-01

Introduction

Affected Products
  1. Persona The Personalized Knee System Tibila Sizing Plate
  2. Persona The Personalized Knee System Natural - Tibia Cemmented 5 Degree Stemmed Provisional

Reason

Potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.