Health Products

Visudep-F Frozen Factors Deficient Plasma (2017-08-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0005-11FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
F. Visudep-F Frozen Factor XI Deficient PlasmaFRDP110105
0004-12FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
G. Visudep-F Frozen Factor XII Deficient PlasmaFRDP120105
0005-8FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
D. Visudep-F Frozen Factor VIII Deficient PlasmaFRDP080125
0004-10FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
E. Visudep-F Frozen Factor X Deficient PlasmaFRDP100105
0004-7FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
B. Visudep-F Frozen Factor VII Deficient PlasmaFRDP070105
0004-9FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
C. Visudep-F Frozen Factor IX Deficient Plasma0004-9FDP
0004-5FDP
Manufacturer
Affinity Biologicals Inc.
1395 Sandhill Drive
Ancaster
L9G 4V5
Ontario
CANADA
A. Visudep-F Frozen Factor V Deficient PlasmaFRDP050105

Introduction

Affected Products
  1. Visudep-F Frozen Factor V Deficient Plasma
  2. Visudep-F Frozen Factor VII Deficient Plasma
  3. Visudep-F Frozen Factor IX Deficient Plasma
  4. Visudep-F Frozen Factor VIII Deficient Plasma
  5. Visudep-F Frozen Factor X Deficient Plasma
  6. Visudep-F Frozen Factor XI Deficient Plasma
  7. Visudep-F Frozen Factor XII Deficient Plasma

Reason

Affinity Biologicals has determined that there was an error in the dilution factors provided in the preparation of the calibration curve section of the IFU (instructions for use) provided with these products.

The provided dilution factors were: 1/10, 1/20, 1/40, 1/180, 1/160.

The correct dilution factors are: 1/10, 1/20, 1/40, 1/80, 1/160.

This is a typographical error only, the description and volumes listed in the table preparation of calibration curve dilutions are correct.