Alaris Pump Module Model 8100 (2017-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerAlaris Pump Module Model 8100
Manufacturer
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
UNITED STATES
8100LVP

Introduction

Affected products

Alaris Pump Module Model 8100

Reason

The separation of one or more posts that connect the mechanism frame to the bezel assembly may prevent the device from delivering an accurate amount of fluid flow through the pumping cycle resulting in an over or under infusion condition. In addition, it may prevent the pump from alarming for an upstream or downstream occlusion condition. The separation of one or more posts are not visible to the end user. An over or under infusion may lead to injuries requiring medical intervention.