Health Products

Brilliance 40/64 Slice and Ingenuity CT Systems (2017-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsB. INGENUITY CT SYSTEM
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
44143
Ohio
UNITED STATES
728323
All lotsA. BRILLIANCE 40/64 SLICE CT SYSTEM
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
44143
Ohio
UNITED STATES
728231

Introduction

Affected products
  1. BRILLIANCE 40/64 SLICE CT SYSTEM
  2. INGENUITY CT SYSTEM

Reason

Upon completion of scan procedure during bolus tracking on a Brilliance iCT with software version 4.1.6, no images were generated. This leads to unavailability of raw data for offline reconstruction.