Health Products

Galileo Neo Immuno Haematology Analyzer (2017-09-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers
Manufacturer
Immucor Inc.
3130 Gateway Drive, P.O. Box 5625
Norcross
30091-5625
Georgia
UNITED STATES
B. Galileo Neo (class 4) Immuno Haematology Analyzer64599
All serial numbers
Manufacturer
Immucor Inc.
3130 Gateway Drive, P.O. Box 5625
Norcross
30091-5625
Georgia
UNITED STATES
A. Galileo Neo Immuno Haematology Analyzer64599

Introduction

Affected products
  1. Galileo Neo Immuno Haematology Analyzer
  2. Galileo Neo (class 4) Immuno Haematology Analyzer

Reason

This recall is to communicate the installation of an update to the Weak D assay on the NEO instrument. This update has been developed to resolve potential invalid results generated when the negative control well (well C1) reacts unexpectedly strongly positive due to droplet accumulation.