B. THERMOCOOL SMARTTOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER
Manufacturer
Biosense Webster Inc.
33 TECHNOLOGY DRIVE
IRVINE
32618
California
UNITED STATES
D134804
17682851L
A. THERMOCOOL SMARTTOUCH SF UNI-DIRECTIONAL NAVIGATION CATHETER
Manufacturer
Biosense Webster Inc.
33 TECHNOLOGY DRIVE
IRVINE
32618
California
UNITED STATES
D134701
Introduction
Affected products
THERMOCOOL SMARTTOUCH SF UNI-DIRECTIONAL NAVIGATION CATHETER
THERMOCOOL SMARTTOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER
Reason
Biosense Webster has received an increased number of complaints related to the display of alert 402 on the Carto® 3 system, for certain lots of THERMOCOOL® brand catheters. Alert 402 implies magnetic distortion which may result in inaccurate catheter tip location information. This may lead the physician to ablate in an unintended area when delivering RF energy. The user manual provides steps for resolving this alert but for the lots subject to this recall, those steps are not effective. Subject lots are therefore being removed.