AUTOCHECK 5+ (2017-07-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Drugs
Issue: Medical Devices
Audience: Healthcare Professionals, Hospitals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
R0689
Manufacturer
Radiometer Medical Aps
Akandevej 21
Bronshoj
2700
DENMARK
AUTOCHECK 5+S7745

Introduction

Affected products

AUTOCHECK 5+

Reason

Radiometer has recently become aware of AUTOCHECK 5+ level 2 ampoules, may give incorrect results for pO2. A defective QC ampoule may for pO2 give a considerable higher result than the expected assigned value of 13.9 - 14.3 kPa (104-107 mmHg).