Health Products

A-SERIES A625 LIFT (2017-09-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, Healthcare Professionals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not ApplicableA-SERIES A625 LIFT
Manufacturer
Prism Medical USA
10888 Metro Ct.
Maryland Heights
63043
Missouri
UNITED STATES
324118

Introduction

Affected products

A-SERIES A625 LIFT

Reason

During internal cycle testing of A-Series lifts Prism Medical Ltd. Has found that the A-Series Lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). They have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.