Health Products

ERGO++ RTP SOFTWARE (2017-09-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
VERSION 1.7.3 AND HIGHER
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems
Inc.
13723 Riverport Drive
Missouri
UNITED STATES
ERGO++ RTP SOFTWARE9600

Introduction

Affected products

ERGO++ RTP SOFTWARE

Reason

Elekta has become aware of the potential for incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.