Health Products

SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6) (2017-10-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6)
Manufacturer
Baxter Healthcare Corporation
711 Park Ave.
Medina
14103
New York
UNITED STATES
35700BAX

Introduction

Affected products

SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6)

Reason

Baxter Corporation is issuing an urgent medical device correction for certain Sigma Spectrum Infusion pumps with Master Drug Library (v6) due to the potential that a small number of these pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication. This could potentially result in over- or under-infusion.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime