Health Products

HEARTSTART MRX DEFIBRILLATOR/MONITOR (2017-09-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
HEARTSTART MRX DEFIBRILLATOR/MONITORM3536A

Introduction

Affected products

HEARTSTART MRX DEFIBRILLATOR/MONITOR

Reason

It was determined that 20 HeartStart MRx Defibrillator/monitors were affected by the same organic contamination clogging the 9 um filter in the zero line of the EtCO2 module. This issue may expose a patient to a hazardous situation because: 
  • A capnogram is not displayed and that may result in unrecognized intubation/ventilation issues 
  • CO2/respiratory rate monitoring is unavailable 
  • EtCO2 will not be displayed, necessitating that the user rely on other parameters to adjust ventilation rate and depth.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime