Architect C16000/C4000 System - Processing Modules (2017-10-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
s/n: C460943
Manufacturer
Abbott Laboratories Diagnostics Division
100 ABBOTT PARK ROAD
ABBOTT PARK
60064
Illinois
UNITED STATES
B. ARCHITECT C4000 SYSTEM - PROCESSING MODULE02P24-01
s/n: C1600144
Manufacturer
Abbott Laboratories Diagnostics Division
100 ABBOTT PARK ROAD
ABBOTT PARK
60064
Illinois
UNITED STATES
A. ARCHITECT C 16000 - PROCESSING MODULE03L77-01

Introduction

Affected products
  1. ARCHITECT C 16000 - PROCESSING MODULE
  2. ARCHITECT C4000 SYSTEM - PROCESSING MODULE

Reason

Abbott has become aware that the Peristaltic Head tubing placed on the Architect has the possibility to leak due to a manufacturing issue associated with the tubing connector. This tubing was not original to Architect, but installed during a service visit that occurred between June 16, 2017 and July 20, 2017. Peristaltic Head tubing is used on Architect clinical chemistry instruments as part of waste aspiration during the cuvette cleaning process.