Health Products

ATLAS GOLD PTA DILATATION CATHETER (2017-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer. ATLAS GOLD PTA DILATATION CATHETER
Manufacturer
Bard Peripheral Vascular Inc.
1625 West 3rd Street
Tempe
55441
UNITED STATES
ATG120162

Introduction

Affected products

ATLAS GOLD PTA DILATATION CATHETER

Reason

To inform customers that specific product code / lot number combinations may be at risk of having a balloon wall thickness that is higher than the required specification, which could result in a slight increase in the force required to advance the balloon through the sheath and/or withdraw the balloon after inflation and deflation, and could potentially lead to a slight prolongation of the procedure. Potentially, there may also be an increase in the balloon's non-compliance and rigidity. Product return is not required.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime