Health Products

Tracoe Twist Tracheostomy Tube size 06 (2017-10-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1000124104Tracoe Twist Tracheostomy Tube size 06
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
306-06

Introduction

Affected products

Tracoe Twist Tracheostomy Tube size 06

Reason

It was established at primary packaging that some obturators of one batch (size 06, lot 1000124104) can break when inserted into the outer cannula.  There is a risk of broken fragments being inhaled into the lower respiratory tract.