Maquet Spring Arms (2017-10-03) Oct 3, 2017 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number up to SN 45060039322 Manufacturer Maquet S.A. Avenue de la Pomme De Pin, Parc De Limere Ardon, Orleans 45074 FRANCE D. XTEN Video Spring Arm ARD567801094 up to SN 44060038859 Manufacturer Maquet S.A. Avenue de la Pomme De Pin, Parc De Limere Ardon, Orleans 45074 FRANCE E. Spring Arm for 3004 TV ARD567910910 up to SN 35060032600 Manufacturer Maquet S.A. Avenue de la Pomme De Pin, Parc De Limere Ardon, Orleans 45074 FRANCE B. Spring Arm for 3000 ARD567910901 up to SN 36060033055 Manufacturer Maquet S.A. Avenue de la Pomme De Pin, Parc De Limere Ardon, Orleans 45074 FRANCE C. XTEN Spring Arm ARD567801093 up to SN 4306038616 Manufacturer Maquet S.A. Avenue de la Pomme De Pin, Parc De Limere Ardon, Orleans 45074 FRANCE Spring Arm for HLX 2000 ARD569002999
Affected products
Spring Arm for HLX 2000
Spring Arm for 3000
XTEN Spring Arm
XTEN Video Spring Arm
Spring Arm for 3004 TV
Fissured or broken weld seams continue to be reported from the field for HLX 2004-5 DF, HLX 3004-5DF and XTEN DF. The spring arms that are still in use are now more than 10 years old. This new field action is launched to replace all remaining spring arms in combination with HLX 2004-5 DF, HLX 3004-5 DF and XTEN DF.